Thermo Scientific
TSQ Altis, Quantis, and Fortis Hardware Manual
103
23
Maintaining the API Source Interface
Clean the ion sweep cone, the spray cone, and the ion transfer tube at the same time, since
they are closely connected.
Perform these tasks in order:
1.
Removing the Ion Sweep Cone and the Ion Transfer Tube
2.
Cleaning the Spray Cone and O-ring
3.
Cleaning the Ion Transfer Tube
4.
Performing Source Maintenance Animation
2. In the search field, type
TSQ Altis
,
TSQ Quantis
, or
TSQ Fortis
.
3. Click the
Catalog
tab and then the instrument name.
4. Scroll down until you see the Product overview, Videos, and Documents tab.
Click the
Videos
tab and on the right, click
Performing source maintenance
.
CAUTION
To prevent corrosion, do not use nitric acid to clean metal parts.
Tip
You do not have to vent the system to remove the ion transfer tube.