![Thermo Scientific Surveyor UV Plus Hardware Manual Download Page 38](http://html1.mh-extra.com/html/thermo-scientific/surveyor-uv-plus/surveyor-uv-plus_hardware-manual_1105375038.webp)
2
Installation
Installing the Flow Cell
22
Surveyor UV/Vis Plus Detector Hardware Manual
Thermo Scientific
• If the autosampler or injector device output relay consists of a single, normally open
terminal and a common ground terminal, connect the normally open terminal to the
RUN terminal on the UV/Vis detector, and connect the common ground terminal to the
GND terminal on the UV/Vis detector. If both of the relay contacts are electrically
isolated from each other and from ground (the device simply closes a contact between the
two relay terminals), then connect either one of the relay terminals to the RUN terminal
on the UV/Vis detector, and connect the other relay terminal to the GND terminal on
the UV/Vis detector.
ZERO Input Signal
The UV/Vis detector does not use the ZERO connection.
EVENT Output Signal
Thermo Scientific data systems do not support this output signal.
Installing the Flow Cell
The Surveyor UV/Vis ships in two configurations (see
When you order the UV/Vis detector with the standard analytical flow cell, the detector
arrives with the flow cell installed. For more information about the standard analytical flow
cell, see
“Standard Analytical Flow Cell”
.
When you order the detector with the LightPipe flow cell, the LightPipe flow cell comes
packed in a small box within the shipping carton for the detector. In addition to the LightPipe
flow cell (with protective caps on each end), this box contains a plastic bag with the inlet and
outlet tubing and FingerTight fittings. The LightPipe mounting assembly (P/N F3005-010)
is connected to the detector’s front panel.
To install the LightPipe flow cell, follow the instructions in
“Installing the LightPipe Flow
.
You can install optional flow cells according to the directions included in their packaging.
lists the optional flow cells for the Surveyor UV/Vis Plus Detector.
Note
The ground (GND) terminals (pins 1 and 7) on the back of the UV/Vis detector are
tied to a single digital ground. You can use either ground terminal for digital input or
output return connections.
Table 4.
Shippable configurations of the Surveyor UV/Vis Plus Detector
Hardware configuration
Part number
10 mm standard analytical flow cell
Surveyor-UV1
50 mm LightPipe flow cell
Surveyor-UV5