Maintenance
LightPipe Flow Cell
Thermo Scientific
Surveyor UV/Vis Plus Detector Hardware Manual
29
4. Carefully pull the photodiode mount straight back and off the LightPipe mounting
assembly. Rest the photodiode mount out of the way inside the detector housing or on
the benchtop.
5. Loosen the thumbscrew holding the retention block to the LightPipe retention bar. See
Figure 21.
Retention block thumbscrew
6. Lift the retention bar shown in
until the bar is clear of the alignment notch in
the LightPipe flow cell.
Figure 22.
Retention bar for the LightPipe flow cell
Thumbscrew
Retaining block
Notch
Retention bar
Optical fiber