Installation
Making the Back Panel Connections
Thermo Scientific
Surveyor UV/Vis Plus Detector Hardware Manual
15
Connecting the UV/Vis Detector to the Data System Computer
Connect your Ethernet switch to the Ethernet connector port of the detector using the
supplied shielded Ethernet cable (P/N 70111-63302).
If your are connecting more than one Surveyor UV/Vis Plus Detector to the same data system
computer, make sure that Unit IDs for the two detectors are set to different values (see
Setting the Unit ID
The UV/Vis detector arrives with the unit ID preset using the two rotary switches located on
the back panel.
shows the rotary switches preset to a value of 01. The range of
values for the unit ID is 01 to 99. The value of 00 is reserved for special service functions.
The unit ID must correspond with the stack ID specified in the Instrument Configuration
application. Do not change the unit ID setting for your detector unless you are controlling
more than one UV/Vis detector from one computer. For information about configuring your
detector, refer to the data system Help.
To change the unit ID setting
Use a small flathead screwdriver to change the setting of the rotary switches.
Figure 11.
Unit ID rotary switches set to a value of 01
1
2 3 4
5
6
7
8
9
0
1
2 3 4
5
6
7
8
9
0