3
Maintenance
LightPipe Flow Cell
Thermo Scientific
Surveyor UV/Vis Plus Detector Hardware Manual
31
Figure 24.
LightPipe mounting assembly shown attached to the front panel
5. Using the 9/64 in. hex driver (supplied), screw the two LightPipe mounting assembly
screws into the detector. Do not overtighten the screws.
Installing the LightPipe Flow Cell
To install the LightPipe flow cell
1. If the photodiode mount is attached to the end of the LightPipe mounting assembly,
detach it as follows.
a. Using the 3/32 in. hex wrench (supplied), loosen and remove the two photodiode
).
b. Carefully pull the photodiode mount straight back and off the LightPipe mounting
assembly. Place the photodiode mount out of the way inside the detector housing or
on the benchtop.
2. Remove the protective end caps from the ends of the LightPipe flow cell.
3. Using the supplied fittings, attach the inlet tubing and the outlet tubing to the flow cell as
follows:
a. Connect the insulated red PEEK inlet tubing to the port in the bottom of the
flow cell.
b. Connect the blue PEEK outlet tubing to the port in the top of the flow cell.
Mounting assembly screws
Retention bar
CAUTION
Do not touch the ends of the LightPipe flow cell. This could damage the
sensitive optics. If you must grasp the ends of the LightPipe flow cell to remove it
from the assembly, use clean, talc-free gloves.