Placement and Environmental Requirements
12
Surveyor UV/Vis Plus Detector Hardware Manual
Thermo Scientific
lists the LightPipe flow cell assembly parts.
Placement and Environmental Requirements
Place the UV/Vis detector on a benchtop as close as possible to the chromatographic
column outlet. This minimizes the length of tubing required to connect the LC column outlet
to the flow cell inlet. If you are using other Surveyor modules, place the UV/Vis detector
on top of the autosampler.
Ensure that the benchtop location meets the following requirements:
• A draft-free location away from an open window, air conditioner vents, or other
circulating air sources.
• A stable room temperature (± 0.5 °C) for applications requiring maximum detection
sensitivity.
• Clearance of at least 15 cm (6 in.) between the back panel of the UV/Vis detector and any
wall or obstruction. This clear space provides access to the back-panel connectors and a
free flow of cooling air.
Installation Tools
Installing the UV/Vis detector requires these tools:
• narrow-tip screwdriver (2 mm wide)
• #2 Phillips screwdriver
• 1/4 in. open-end wrench
• 1/8 in. hex wrench (supplied)
• 3/32 in. hex wrench (supplied)
• 9/64 in. hex wrench (supplied)
Table 3.
Parts for LightPipe flow cell assembly
Description
Part number
Flow cell assembly, with inlet/outlet tubing and fittings
(5 cm LightPipe)
803237
FingerTight PEEK™ ferrule nuts
2522-0285
Inlet tubing, with insulation, PEEK 1/16 × 0.005 in. ID (red)
803260
Outlet tubing, PEEK 1/16 × 0.010 in. ID (blue)
703950