![Thermo Scientific SOLA II Flare User Manual Download Page 76](http://html1.mh-extra.com/html/thermo-scientific/sola-ii-flare/sola-ii-flare_user-manual_1105273076.webp)
Chapter 7
Diagnostics
View Digital Inputs
When you access the View Digital Inputs menu, the status of the inputs is
reported but the functions are ignored.
I n p u t
&
O u t p u t
T e s t
V i e w
D i g i t a l
I n p u t s
∨
∧
E n t
E x i t
Figure 7-6.
Following is a list of inputs and their functions.
Table 7-2.
Input
Function
Purge Fail
Displays the status of the purge sense pressure switch.
Flow Switch
Displays the status of the flow monitoring switch.
Watlow Alarm
Displays the status of the Watlow alarm input
I/O Board
Displays the status of all eight inputs simultaneously. Zero (0) denotes open.
One (1) denotes closed.
CPU Digital In
Dip Switches
Displays the positions of dip switches located in the motherboard (dip
switches are used to indicate options installed)
I/O Board Option
4–20 mA Input Option
Not used in SOLA II Flare
Com 4 Modem
Modbus input
Remote Suspend
Displays the status of the remote suspend switch
View Parameters
Each menu item within the View Parameters menu displays the parameter
name, current value, minimum value, and maximum value. Displaying the
minimum, maximum, and current values for a period of time enables you to
monitor signal stability.
Note
When you select a parameter to view, the minimum and maximum values
are briefly reset to the current value.
▲
Thermo Scientific
SOLA II Flare User
Manual
7-4
Summary of Contents for SOLA II Flare
Page 1: ...SOLA II Flare Sulfur Online Analyzer User Manual PN 90 1312 0...
Page 2: ......
Page 3: ...SOLA II Flare Sulfur Online Analyzer User Manual PN 90 1312 0 Revision A...
Page 4: ......
Page 6: ......
Page 8: ......
Page 14: ......
Page 18: ......
Page 44: ......
Page 48: ......
Page 70: ......
Page 72: ......
Page 80: ......
Page 104: ......
Page 122: ......
Page 136: ......
Page 148: ......
Page 158: ......
Page 160: ......
Page 162: ......