Chapter 2
Installation
Thermo Scientific
SOLA II Flare User Manual
2-6
Installation Checklist
The following checklist may be copied for use during system installation.
Details for the installation and various specification requirements are contained
in this manual, application specific drawings, and information supplied with
the instrument.
Materials used meet specifications defined in
Operating environment meets requirements defined in
Mounting meets requirements defined in
Analyzer condition inspected as follows:
No physical damage, broken parts, or observable defects.
All electronic boards are securely seated.
All cables and wiring connectors are in place and fully seated
No loose parts (wires, nuts, screws, cables, debris, etc.).
All tubing is properly connected and fittings are tight.
Installation and flow connections adhere to the following:
Sample system properly designed to condition (control pressure, flow,
temperature, and particulate) sample as required by system.
Sample system fast loop panel is installed as close as possible to analyzer
for fastest response time.
System gases meet quality and quantity (pressure and flow)
specifications according to
PUVF vent is vented to atmospheric pressure.
PUVF vent is NOT connected to headers with varying pressure or that
may accumulate liquids.
All tubing supply lines to analyzer are sized to match or exceed
connector size on the analyzer.
All internal and external tubing fittings are physically checked for
tightness and tested for leaks.
Installation and electrical connections adhere to the following:
AC power wiring meets requirements.
AC power wiring is properly connected to the instrument with proper
earth grounding.
Signal wiring (dc signals, communications, etc.) meets requirements.
Signal wires properly connected to the instrument.
All electrical conduit seals poured with sealing compound such as Chico
A5r or equivalent.
Summary of Contents for SOLA II Flare
Page 1: ...SOLA II Flare Sulfur Online Analyzer User Manual PN 90 1312 0...
Page 2: ......
Page 3: ...SOLA II Flare Sulfur Online Analyzer User Manual PN 90 1312 0 Revision A...
Page 4: ......
Page 6: ......
Page 8: ......
Page 14: ......
Page 18: ......
Page 44: ......
Page 48: ......
Page 70: ......
Page 72: ......
Page 80: ......
Page 104: ......
Page 122: ......
Page 136: ......
Page 148: ......
Page 158: ......
Page 160: ......
Page 162: ......