Chapter 4
Operation
Thermo Scientific
SOLA II Flare User Manual
4-2
3.
Open sample flow to the analyzer.
4.
Apply power to the instrument.
5.
Analysis begins automatically when the pyrolyzer and oven reach
operating temperature. If you need to change the configuration, press
any button, and enter the passcode.
Suspending an
Analysis
Follow these steps to suspend an analysis.
1.
Enter the Analyzer Mode top-level menu.
2.
Select Suspended to suspend the analysis. In suspended mode, the
analyzer stops injecting, and shuts off the air to the sampling system
pneumatic and allows only air to run through the burner and mixing
chamber.
3.
To resume the analysis, go to the Analyzer Mode menu, and select
Active.
A n a l y
z
e
r
M
o
d
e
A c t i v
e
E x
i t
Figure 4-2.
Injection Valve
Failure Alarm
Over time, the injection valves begin to wear, resulting in an increased potential
for port-to-port leakage of sample. Such leakage can cause erroneous
measurement and materials buildup in the system. When the analyzer detects
sample leakage through the injection valve, it triggers the Injection Valve
Failure alarm and stops airflow to the pneumatic valves.
Note
The Injection Valve Failure alarm must be acknowledged after valve repair
or replacement.
▲
Toggle
Summary of Contents for SOLA II Flare
Page 1: ...SOLA II Flare Sulfur Online Analyzer User Manual PN 90 1312 0...
Page 2: ......
Page 3: ...SOLA II Flare Sulfur Online Analyzer User Manual PN 90 1312 0 Revision A...
Page 4: ......
Page 6: ......
Page 8: ......
Page 14: ......
Page 18: ......
Page 44: ......
Page 48: ......
Page 70: ......
Page 72: ......
Page 80: ......
Page 104: ......
Page 122: ......
Page 136: ......
Page 148: ......
Page 158: ......
Page 160: ......
Page 162: ......