Appendix C: TCP/IP Bridge
The SOLA II Flare Modbus TCP/IP option is a microprocessor-driven board
that provides a bridge between the SOLA II Flare standard serial Modbus and a
TCP/IP Modbus network using 10/100 Base-T. This function is provided by
the NetBurner™.
Hardware
Configuration
•
TCP/IP Bridge board installed inside the SOLA II Flare enclosure
•
SOLA II Flare Modbus configuration must be set to 38400 baud with
Modbus ID 1
•
Bridge board has a standard RJ45 connector for the 10/100 Base-T
cable
Figure C–1. TCP/IP Ethernet bridge board, P/N 55-1228-0 (newer models)
Software
Configuration
As shipped, the Bridge defaults to an IP address of 00.00.00.00. At power- up
(with an IP address of 00.00.00.00), the Bridge attempts to negotiate an IP
through a DHCP server. If a DHCP server is not available or if it is desired to
assign the instrument a static IP, use the provided utility program to set up the
IP address. Perform the following steps to search for the Bridge on the network
and assign an IP address.
1. Locate the IPSetup.exe utility that is included with the disk that ships
with the TCP/IP Bridge option.
Thermo Scientific
SOLA II Flare User
Manual
C-1
Summary of Contents for SOLA II Flare
Page 1: ...SOLA II Flare Sulfur Online Analyzer User Manual PN 90 1312 0...
Page 2: ......
Page 3: ...SOLA II Flare Sulfur Online Analyzer User Manual PN 90 1312 0 Revision A...
Page 4: ......
Page 6: ......
Page 8: ......
Page 14: ......
Page 18: ......
Page 44: ......
Page 48: ......
Page 70: ......
Page 72: ......
Page 80: ......
Page 104: ......
Page 122: ......
Page 136: ......
Page 148: ......
Page 158: ......
Page 160: ......
Page 162: ......