Chapter 1
Product Overview
Thermo Scientific
SOLA II Flare User Manual
1-11
System
Requirements
Table 1-4.
Ambient
Temperature
54 to 104°F (12 – 40°C) Ingress Protection: IP-51 / dustproof and dripping
water
Power
110 Vac (optional 220 Vac), 50/60 Hz at 2000 watts /±10 volts
20-amp circuit recommended; 17 amps maximum during warm-up cycle; 7-8
amps once achieving operational temperatures
Available
Certifications
CSA/NRTL Class 1, Division 2, Group B, C & D (standard) CSA/NRTL Class 1,
Division 1, Group B, C & D (optional)
System Configuration
Dimensions:
44”H x 26”W x 18"D
1118H x 660W x 457D mm
Weight:
250 lbs (estimated)
114 kg (estimated)
Site Requirements
The customer is responsible for installation of the analyzer system including all
sample, vent, utility electrical and signal connections. The installation should
be completed prior to a Thermo Scientific service technician’s arrival at the
analyzer location.
Table 1-5.
Analyzer Sample
Vent
No backpressure may be applied and the PUVF Detector sample vent – the
vented gases will contain CO
2
, H
2
O and high levels of SO
2
.
Sample
The suggested sample input pressure is dependent upon the type of sample
and therefore is application dependent. For most flare applications the
sample delivered to the SOLA II Flare needs to be in the 25-30 psig range.
Carrier & Auxiliary
Air
Air Carrier Gas (Zero grade air)
Minimum input pressure is 80 psig. The air used as a carrier gas must be
clean, dry and sulfur free. Cylinder air is recommended to obtain this quality.
Consumption is 250-300 cc/min.
Instrument Air
(Operational)
Minimum input pressure 60 psig. The instrument air must be clean, dry and
oil free. Instrument air is required for both the cabinet purge and the μl
injection valve. Consumption is 6-7 SCFM.
Nitrogen (Sweep
Gas)
Minimum input pressure 60 psig. The nitrogen must be clean, dry and oil
free. nitrogen is required for sweeping the unused valve only. Consumption –
when required.
Power
120 Vac - 50/60 Hz
The power supply to the analyzer system should be stable within ±10 volts
Summary of Contents for SOLA II Flare
Page 1: ...SOLA II Flare Sulfur Online Analyzer User Manual PN 90 1312 0...
Page 2: ......
Page 3: ...SOLA II Flare Sulfur Online Analyzer User Manual PN 90 1312 0 Revision A...
Page 4: ......
Page 6: ......
Page 8: ......
Page 14: ......
Page 18: ......
Page 44: ......
Page 48: ......
Page 70: ......
Page 72: ......
Page 80: ......
Page 104: ......
Page 122: ......
Page 136: ......
Page 148: ......
Page 158: ......
Page 160: ......
Page 162: ......