12
A79210100 - Iss 2
Shandon ClearVue™ Operator Guide
DRAFT
Dimensions
height
500 mm
19.7”
Width
645 mm
25.4”
Depth
575 mm
22.6”
Weight
48 kg
106 lbs
Electrical Specification
Voltage
100 - 240 V
ac
Frequency
50 / 60 hz
Power (max)
300 VA
Earth Leakage
< 500
m
A at 110-120 V
ac
Fuses
T 6.3 A, 250 V
Internal batteries
Sealed Lead Acid type.
Not user replaceable.
Environmental Requirements
Warning - For Indoor Use Only
Temperature
(Operating Limits)
+5°C to +40°C (+41°F to +104°F)
Temperature
(Recommended
Operation)
+15°C to +30°C (+59°F to +86°F)
Note: Performance may deteriorate when
operated outside of this temperature range.
Temperature
(Storage)
-25°C to +55°C (-13°F to 131°F)
+70°C (158°F) for short exposure
Relative humidity
Max. 80% Rh up to 31°C
Decreasing linearly to 50% Rh at 40°C
Altitude
Up to 2000 m (6,500 ft)
Pollution Degree
2
Over Voltage
Category
II
1-5 - System
Specification
1-5 - System
Specification
Summary of Contents for Shandon ClearVue A79210100
Page 96: ...96 A79210100 Iss 2 Shandon ClearVue Operator Guide DRAFT 5 2 2 Mountant Jar...
Page 119: ...Shandon ClearVue Operator Guide A79210100 Iss 2 119 DRAFT C2 Packing Instructions...
Page 120: ...120 A79210100 Iss 2 Shandon ClearVue Operator Guide DRAFT Appendix D Screen Maps A B...
Page 121: ...Shandon ClearVue Operator Guide A79210100 Iss 2 121 DRAFT A 1 3 2 4 5 1 2 4 3 5 A...
Page 122: ...122 A79210100 Iss 2 Shandon ClearVue Operator Guide DRAFT B 1 3 2 4 5 7 6 8 1 3 5 7 2 4 6 8...
Page 123: ...Shandon ClearVue Operator Guide A79210100 Iss 2 123 DRAFT Appendix E Declaration of Conformity...
Page 124: ...124 A79210100 Iss 2 Shandon ClearVue Operator Guide DRAFT Index...
Page 125: ...Shandon ClearVue Operator Guide A79210100 Iss 2 125 DRAFT...