Shandon ClearVue™ Operator Guide
A79210100 - Iss 2
85
DRAFT
Problem
Cause
Remedy
basket Jams.
Incorrect size
of slide.
Refer to
Section
1-2
for maximum
permissible
slide sizes.
basket slider not
fully closed.
Ensure slider is
fully closed prior
to loading basket.
Damaged or
broken basket.
Replace basket.
build-up of
dried mountant
on basket.
Clean or replace
basket.
Other error.
Call Thermo Fisher
Service Team
for support.
Summary of Contents for Shandon ClearVue A79210100
Page 96: ...96 A79210100 Iss 2 Shandon ClearVue Operator Guide DRAFT 5 2 2 Mountant Jar...
Page 119: ...Shandon ClearVue Operator Guide A79210100 Iss 2 119 DRAFT C2 Packing Instructions...
Page 120: ...120 A79210100 Iss 2 Shandon ClearVue Operator Guide DRAFT Appendix D Screen Maps A B...
Page 121: ...Shandon ClearVue Operator Guide A79210100 Iss 2 121 DRAFT A 1 3 2 4 5 1 2 4 3 5 A...
Page 122: ...122 A79210100 Iss 2 Shandon ClearVue Operator Guide DRAFT B 1 3 2 4 5 7 6 8 1 3 5 7 2 4 6 8...
Page 123: ...Shandon ClearVue Operator Guide A79210100 Iss 2 123 DRAFT Appendix E Declaration of Conformity...
Page 124: ...124 A79210100 Iss 2 Shandon ClearVue Operator Guide DRAFT Index...
Page 125: ...Shandon ClearVue Operator Guide A79210100 Iss 2 125 DRAFT...