Shandon ClearVue™ Operator Guide
A79210100 - Iss 2
117
DRAFT
Appendix C - Transportation Instructions
C1 - Things to Do Before Packing
Ensure the instructions for cleaning (see
Section
5-3
) have been followed.
Discard chemicals according to local environmental procedures.
Remove the following:
Debris Tray
Remove any discarded glass
Clean any mountant off
Dry
Dispense Head Cleaning Station
Discard any liquid
Clean with xylene
Dry
Xylene tray
Discard any liquid
Dry
Hopper
Remove the
Dispense Head
(see
Section 5-2-8
)
and place in a suitable container to allow it to
drain.
Remove the
Mountant Bottle
and rinse with
xylene, or replace with a fresh bottle.
Pour 25ml of xylene into the clean
Mountant
Bottle
and
Flush
the system (see
Section 2-10
).
Replace the
Dispense Head.
•
•
•
•
•
•
•
•
•
•
•
•
•
•
•
•
•
•