Shandon ClearVue™ Operator Guide
A79210100 - Iss 2
77
DRAFT
Description
Possible Cause
Solution
Single Bubble at Non-
Frosted End of Slide
Often located underneath
the
Suction Cup
.
•
Dispense Head
too high.
Check
Dispense Head
is
fully screwed down.
Contact Thermo Fisher
Scientific Service
Department.
•
•
Single Bubble at Frosted
End of Slide
Dispense Head
too high.
Check
Dispense Head
is
fully screwed down.
Contact Thermo Fisher
Scientific Service
Department.
•
•
Air in
Pipes
or
Syringe
.
Purge
system until all bubbles
have been expelled.
Coverslip contacting the
mountant before the
Transfer
Head
has slowed down.
Contact Thermo Fisher Scientific
Service Department.
Non-approved mountant used.
Use approved mountant (see
Appendix B
)
Distinct Area with No
Mountant
Area of coverslip is not
stuck to slide.
•
Debris under the coverslip.
Ensure coverslips are free
from debris and dust.
Ensure
Slip Dispense Area
is free of debris and dust.
•
•
Very low volume of mountant,
AND slide is very dry.
Ensure an appropriate
amount of mountant is being
dispensed.
Ensure
Xylene Tray
is filled.
Ensure slides are
appropriately coated with
xylene.
•
•
•
Summary of Contents for Shandon ClearVue A79210100
Page 96: ...96 A79210100 Iss 2 Shandon ClearVue Operator Guide DRAFT 5 2 2 Mountant Jar...
Page 119: ...Shandon ClearVue Operator Guide A79210100 Iss 2 119 DRAFT C2 Packing Instructions...
Page 120: ...120 A79210100 Iss 2 Shandon ClearVue Operator Guide DRAFT Appendix D Screen Maps A B...
Page 121: ...Shandon ClearVue Operator Guide A79210100 Iss 2 121 DRAFT A 1 3 2 4 5 1 2 4 3 5 A...
Page 122: ...122 A79210100 Iss 2 Shandon ClearVue Operator Guide DRAFT B 1 3 2 4 5 7 6 8 1 3 5 7 2 4 6 8...
Page 123: ...Shandon ClearVue Operator Guide A79210100 Iss 2 123 DRAFT Appendix E Declaration of Conformity...
Page 124: ...124 A79210100 Iss 2 Shandon ClearVue Operator Guide DRAFT Index...
Page 125: ...Shandon ClearVue Operator Guide A79210100 Iss 2 125 DRAFT...