CHAPTER 3
HYBRIDISATION GUIDE
HYBRIDISATION PROCEDURES
Nucleic acid hybridisations are conveniently performed in the dedicated Hybridisation
equipment available from Thermo.
The Hybridisation Oven range consists of incubators with integral rotisserie devices,
which can accommodate 4, 10 or 14 Hybridisation bottles, 35mm in diameter, in order to
perform Hybridisations in minimal probe volumes with fluid moving continually over the
membrane.
Southern Blot DNA Hybridisations
The following protocol is broadly applicable to Hybridisations using DNA probes of 50bp
and above, following transfer and fixing of DNA to membranes as described in Chapter
2. For Hybridisations using oligonucleotides refer to Chapter 6. The Hybridisation
procedure consists of four stages: -
1. Prehybridisation
2. Hybridisation
3. Stringency
washing
4. Autoradiography
For detailed notes on Hybridisation specific to Thermo equipment, refer to ‘Notes for
Nucleic Acid Hybridisation’ on page 11.
1.
Prehybridisation
is carried out by incubating the membrane in
Southern Blot
Prehybridisation Buffer
(for recipe see page 30). Denature salmon sperm DNA
by boiling for 5 minutes and then chilling on ice. Add the denatured salmon sperm
DNA to the buffer to a final concentration of 50
µ
g/ml.
2.
Incubate with agitation or in a rotisserie for a least 1-hour at 65
°
C.
3.
The volume of prehybridisation buffer required varies according to the Hybridisation
system being utilised. In general terms, the minimum volume of buffer should be
used such that the membrane is covered by the fluid at all times (approximately
0.1ml/cm
2
), or if in Hybridisation bottles, 10ml for a large bottle and 5ml for a small
bottle.
Summary of Contents for Shake 'n Stack 6244
Page 13: ...2 4 Shake n Stack Thermo Scientific Section 2 Unpacking and Installation...
Page 21: ...4 6 Shake n Stack Thermo Scientific Section 4 Methodology of Hybridisation...
Page 27: ...7 2 Shake n Stack Thermo Scientific Section 7 Use of Radioactive Probes...
Page 31: ...8 4 Shake n Stack Thermo Scientific Section 8 Technical Specifications...
Page 32: ...Shake n Stack 8 5 Thermo Scientific Section 8 Technical Specifications...
Page 33: ...8 6 Shake n Stack Thermo Scientific Section 8 Technical Specifications...
Page 34: ...Shake n Stack 8 7 Thermo Scientific Section 8 Technical Specifications...
Page 35: ...8 8 Shake n Stack Thermo Scientific Section 8 Technical Specifications...
Page 36: ...Shake n Stack 8 9 Thermo Scientific Section 8 Technical Specifications...
Page 37: ...8 10 Shake n Stack Thermo Scientific Section 8 Technical Specifications...
Page 38: ...Shake n Stack 9 1 Thermo Scientific Section 9 Electrical Schematics...
Page 39: ...9 2 Shake n Stack Thermo Scientific Section 9 Electrical Schematics...
Page 40: ...Shake n Stack 9 3 Thermo Scientific Section 9 Electrical Schematics...
Page 41: ...9 4 Shake n Stack Thermo Scientific Section 9 Electrical Schematics...
Page 42: ......
Page 50: ...A 8 Shake n Stack Thermo Scientific...
Page 51: ...HYBRIDISATION GUIDE USER INSTRUCTION MANUAL Manual 7222060 Rev 0...
Page 91: ......