Terms Definition
Section 8
2230XP Silica Analyzer
Thermo Scientific Orion
61
Terms
Description
Possible Values, Options
Display Help
Help message for the user
to navigate between the menus.
Enabled, Disabled
Display Info
Message showing the status
of the current cycle performing.
Enabled, Disabled
Results Decimal Point
Selectable precision to show the
analysis result.
XXXX.X, XXX.XX, XX.XXX, X.XXX,
XXXXX, Auto
Large Fonts Screen
To display the result with large
fonts.
Enabled, Disabled
CALIBRATION PARAMETERS
Standard
The concentration of the standard
used for calibration.
0 to XXXXX (XXXXX Value depends on
precision selected in Setup
➔
Display
parameters
➔
result Decimal Point
Mode
Mode represents the calibration
type.
Single-Point, Base-Line
Frequency
The time duration between two
calibration cycles.
Every 6Hr, 12Hr, 24Hr, Sun, Mon, Tue,
wed, Thu, Fri, Sat, Manual and Editable
0Hr to 99.99Hr
Start Time
Calibrations start time, except for
manual and editable frequency.
Depends on Time format in Setup
➔
Time Menu
Slope
The CURRENT Slope value
represents the coefficient that is
multiplied by the absorbance to
calculate the concentration of the
/- Intercept value.
Intercept
Reference and current numbers
based on the base line calibration
cycle.
Default value of -1
Calibration Tolerance
Repeat tolerance tested during auto
calibration cycles to accept or reject
automatic calibrations.
Default value of +/- 5%
VALIDATION PARAMETERS
Validation Source
Fluid source for validation.
Sample, Standard, Grab Sample, Not
Assigned
Number of Repeats
Number of times that the validation
is to be performed.
0 to 8
ANALYSIS PARAMETERS
Mode
Instrument RESTART mode.
Auto Restart, Manual Restart
Measuring Frequency
In the RUN mode, the frequency
is the time duration between two
analysis cycles.
15, 30, 45 minutes, 1 hr, 0.20 hr are
selectable and editable minutes range
00.00 to 99.99Hr
Measuring Units
Units for the analyzed sample.
Six characters long. Alphabets,
numerals and special symbols are
allowed.
Analysis Label
Six characters long. Alphabets,
Numerals and +, - are allowed.
Cell Wash Flow Rate
The rate at which the sample
is pumped to wash the cell.
Depends on the inner diameter of the
pump tubing. It is derived from the
calibration of pumps.