Introduction
Section 2
2230XP Silica Analyzer
Thermo Scientific Orion
11
b.
Select
Services.
c.
Select
Prime Fluidics.
d.
Select
Prime All.
Note:
This procedure will run automatically and take approximately 10 minutes per
cycle.
5.
Refer to Section 4 for more details on startup and operation. Once the above 4 steps have
been completed, (4. a, b, c, d) press
Stop
,
Select Run
. Allow the unit to operate overnight
or for at least 12 hours, to become stabilized in the local environment.
6.
For additional information on 2230XP Orion Silica Process Analyzer Operation, refer to
Section 4.
Additional detail can be found in the following sections:
Section 4 - Instrument Operation
Section 5 - Analyzer Hardware Configuration and Calibration
Section 6 - Calibration of the Instrument for Measurement Analysis
Section 7 - Menu Overview
Section 8 - Terms Definition
Section 9 - Maintenance Schedule
General Product Description
This User Manual provides the information necessary to obtain the maximum benefits from the
Orion 2230XP Silica Analyzer. All operating instructions, functional descriptions, illustrations,
and other relevant information are contained in this manual. Descriptions of the system
hardware and user-interface software (the control panel) are included, as is an overview of the
Image: Reagent bottle numbers in
proper order