Maintenance Schedule
Section 9
2230XP Silica Analyzer
Thermo Scientific Orion
73
Messages
Cause
Solution
No records to display
User is trying to display records, and
no records are available in the
selected data range.
None.
No records to print
User is trying to print records, and no
records are available in the selected
data range.
None.
Only one record to display
User is trying to view graph when only
one analysis record is available in the
database.
None.
Problem with memory
allocation. Insufficient
memory in the system
or system is busy.
Do not try to repeatedly view the
graph.
Try again after some duration or
after performing one more
measurement cycle.
Measurement/Calibration/Validation Messages
Messages
Cause
Solution
Reagent1 empty
Performing Analysis/ Calibration/
Validation with two or three Reagents
and Reagent1 is exhausted.
Fill the empty Reagent1 bottle
with Reagent1.
Reagent2 empty
Performing Analysis/ Calibration/
Validation with two or three Reagents
and Reagent2 is exhausted.
Fill the empty Reagent2 bottle
with Reagent2.
Reagent3 empty
Performing Analysis/ Calibration/
Validation with two or three Reagents
and Reagent3 is exhausted.
Fill the empty reagent3 bottle
with Reagent3.
Reagent1, Reagent2 are
empty
Performing Analysis/ Calibration/
Validation with two or three Reagents
and Reagent1 & 2 are exhausted.
Fill the empty Reagent1 & 2
bottles with Reagent1 & 2.
Standard empty
Performing single point or baseline
calibration and the Standard is
exhausted.
Fill the empty standard bottle with
standard.
Low standard empty
Performing two-point calibration
and the Low Standard is exhausted.
Fill the empty Low-Standard
bottle with Low-Standard.
Hi standard empty
Performing two-point calibration
and the Hi Standard is exhausted.
Fill the empty Hi-Standard bottle
with Hi-Standard.
Two standards are empty
Performing two-point calibration
and both standards are exhausted.
Fill the empty standard bottles
with their respective standards.
Sample empty
Performing analysis with single
sample and sample chamber is empty.
Check for Sample flow.
Check for pipe blockages
to sample chamber.
Testing optics, please check
for optics failure
No light output from optics.
Run a Clean Cycle and try again.
Inspect cell and clean, check for
light detection under LED &
DETECTOR CONTROL in
Diagnostics menu.