2230XP Silica Analyzer
Thermo Scientific Orion
17
3
Installation Environmental
Considerations
The instrument is designed for general-duty, indoor installations. The electronics enclosure
environmental rating is NEMA 12 and the Fluidics enclosure is a NEMA 4X fiberglass with a
clear door window.
The Orion 2230XP Silica Analyzer should be installed in a well-ventilated location, with ambient
temperature between 5°C and 45°C (41° to 113°F); and relative humidity not to exceed 90% at
40°C. The area must be free from dust, corrosive gases, vibration and shocks; sheltered from
direct sunlight, and shielded from dripping water. Power should be instrument grade with over
voltage protection, and a provision for disconnecting power.
Mounting the Instrument
The instrument case is designed for wall or panel mounting.
Use four #10 screws or equivalent for proper support.
Refer to Figure 2 and Figure 3 for dimensions (in mm and inches).
Make sure that the wall that you choose can support the weight of the instrument
(approximately 20 kg or 44 lbs.) See Figure 2.
Locate the instrument as close to the sampling point as practical to ensure complete purging
of the sample line during each cycle.
Allow approximately 50 cm (20 in.) of clearance at the left side for door swing. Also allow
enough clearance on the right side for wiring conduit.
Locate the analyzer height such that the display is at eye-level height for an operator,
allowing also for appropriate sample and drain connections.
S E C T I O N 3