Section 1
General
Information
8
Thermo Scientific Orion
2230XP Silica Analyzer
Electrical
Power Requirements
110 watts. Automatic voltage recognition from 100 – 240 VAC, 110 w (VA), 50 – 60 Hz,
Single Phase
Connection
Three wire terminal block (18 – 12 AwG) through a 19 mm (3/4-in) bulkhead on side of
the electronics enclosure
Safety Class
I, protective grounding
Fuse Rating
250 VAC, 4 A (fast blow)
Relay Contacts
Four potential free relays with 250 VAC, 2A resistive contact ratings.
Outputs
Each relay may be configured separately to a list of programmable alarm functions
Inputs
Two-isolated analog outputs: 0/4-20mA, max. load 900 ohms
Environmental
Storage Temperature Range
5 – 45°C (41 – 113°F)
Operating Temperature Range
5 – 45 °C (41 – 113°F)
Relative Humidity
90% at 40°C (104°F), non-condensing
Altitude
Up to 2000 m
Sample Requirements
Inlet Pressure to Instrument
1 – 5 psig
Sample Flow Rate
50 – 1000 mL/min with < 100 microns particles size (user provides filter)
Sample Temperature
5 – 40°C (41 – 104°F)
Sample Fitting
6 mm (1/4-in) OD tubing
Drain Fitting
9.6 mm (3/8-in) OD tubing