2230XP Silica Analyzer
Thermo Scientific Orion
66
9
Maintenance Schedule
Maintenance
Schedule Frequency
Conduct regular maintenance
to ensure the analyzer performs properly.
Every week
Inspect in / out tubes, reagent bottle, and all tubes for fouling.
Every month
Inspect tubing and ensure that it is not blocked.
Every 6 months
Replace reagents if necessary.
Every 12 months
Replace Pump and valve tubing. Use factory supplied maintenance kit. Inspect
for any fouling of the reaction cell.
Changing Reagents
CAUTION
Wear gloves or protective equipment your site may require when
changing reagents .
Press the
STOP
key.
Press the
MENU
key.
Remove the cap from each bottle and change them over one by one. Do
NOT re-use or re-fill old or existing bottles. Dispose of the used bottles and
contents safely and according to local regulations.
Tighten bottle caps and route the tubing properly making sure that it does
not get pinched when placing the bottle in the enclosure.
Press
MENU
key and select:
SERVICES
➔
PRIME FLUIDICS
➔
AUTO PRIME
Repeat AUTO PRIME if you notice air bubbles in the tubing. Each reagents
kit comes with a REAGENT BLANK value. After changing the reagent you
must enter the new value.
S E C T I O N 9