Section 7
Menu
Overview
54
Thermo Scientific Orion
2230XP Silica Analyzer
Factory Settings Menu
Allows technician to restore factory defaults, update factory defaults and edit factory settings.
Instrument Details Menu
Allows viewing of instrument details:
Software version
Serial number
Model number
Analysis Label Change Menu
Allows technician to edit the analysis label.
Printing Options Menu
Allows technician to print the system configuration, analysis parameters and method
parameters. Also allows technician to enable/disable online printing. See Appendix A for the
default analyzer parameters listing printouts.
Hardware Setup Menu
Calibrate Thermistor
Allows technician to calibrate the thermistor. Only required if the factory-installed thermistor has
been serviced or replaced. To calibrate the thermistor, user is required to enter the reaction
block temperature using an external reference thermometer.
Calibrate Pumps
Allows technician to calibrate the flow rate of the pumps.
Calibrate Optics
Allows technician to calibrate the optics.
Calibrate Analog Outputs
Allows technician to calibrate the two analog outputs.
Edit Calibrated Values
Allows technician to edit the calibrated results for fine-tuning.