Error Codes
Thermo Scientific
Heratherm
12-3
Calibration value too
low (E105)
The calibration reference
value calculated on the
basis of the user input
exceeds the lower limit for
calibration references.
Fall back to previous cali-
bration reference. Audible
alarm activated, alarm
relay energized, message
shown on display.
Check the external refer-
ence sensor for proper
function and replace, if
necessary. If this doesn’t
solve the problem, call
service.
Constant sensor signal
(E106)
None of the decimal
places of the A/D-con-
verter output for the pro-
cess sensor has changed
over a specific time
period.
Control transferred to ref-
erence sensor. Audible
alarm activated, alarm
relay energized, message
shown on display. If both
sensors are defective, all
control circuits will be dis-
abled.
Call service.
Constant reference
sensor signal (E107)
None of the decimal
places of the A/D-con-
verter output for the pro-
cess sensor has changed
over a specific time
period.
Process control continues
on set value. Audible
alarm activated, alarm
relay energized, message
shown on display. If both
sensors are defective, all
control circuits will be dis-
abled.
Call service.
Heating relay error
(E109)
The voltage measure-
ment has indicated a
defect in the heater circuit
relay.
Device completely inoper-
ative, audible alarm acti-
vated, alarm relay
energized, message
shown on display.
Call service.
Heating triac error
(E110)
The voltage measure-
ment has indicated a
defect in the triac.
Audible alarm activated,
alarm relay energized,
message shown on dis-
play. Audible alarm
returns upon acknowl-
edgement.
Switch off the oven and
call service.
Temperature too high
(E111)
Actual measured value
exceeds permissible
range. May occur when
the user lowers the set
point temperature on a hot
device whose door is
closed.
Audible alarm activated,
alarm relay energized,
message shown on dis-
play. Heater turned off
until upper limit of hystere-
sis is recovered. Servo
control operations con-
tinue. Alarm can be
acknowledged, and goes
away when the difference
between the actual and
set values ceases to exist.
Note: This error does not
indicate a defective triac.
Open the door to speed
up cool-down. Check
whether the device was
loaded with a hot object, if
so, remove. Ensure that
the equipment was oper-
ated with at least one per-
forated shelve and with
the door not opened lon-
ger than 10 min. If this
doesn’t solve the problem,
call service.
For 3P-units only:
when set-point tempera-
ture is stabilized, switch
unit to Standby Mode and
turn-on again. If unit can-
not keep temperature
without failure please call
Technical Support.
Table 12-1 Heratherm Oven Error Codes
Error Message & Code
Root Cause
Alarm Response
Alarm Clearing Instructions
1
Summary of Contents for OGS 100
Page 1: ......
Page 6: ...Contents iv Heratherm Thermo Scientific...
Page 46: ...Product Description Tube Access Ports 4 16 Heratherm Thermo Scientific...
Page 60: ...Start up Connecting the RS 232 Interface 5 14 Heratherm Thermo Scientific...
Page 76: ...Handling and Control Switching the Oven Off Powering Down 7 14 Heratherm Thermo Scientific...
Page 78: ...Shut down Shutting the Oven Down 8 2 Heratherm Thermo Scientific...
Page 82: ...Cleaning and Disinfection Wipe Spray Disinfection 9 4 Heratherm Thermo Scientific...
Page 88: ...Maintenance Returns for Repair 10 6 Heratherm Thermo Scientific...
Page 104: ...Spare Parts and Accessories 14 6 Heratherm Thermo Scientific...
Page 106: ...Device Log 15 2 Heratherm Thermo Scientific...
Page 108: ...Contact 16 2 Heratherm Thermo Scientific...