Operating Environment
Air Conditioning
3-6
iCAP RQ Pre-Installation Requirements Guide (P/N BRE0009927, Revision A)
Thermo Scientific
Air Conditioning
The air conditioning system must be capable of maintaining a constant
temperature as specified in
in the immediate
vicinity of the system.
Thermo Fisher Scientific recommends the installation of an air
conditioner if the specified limits are exceeded due to unfavorable
climatic conditions. For reliable operation of the iCAP RQ instrument,
a stability of the room temperature < 2 °C/h is required.
NOTICE
Do not locate the iCAP RQ instrument under an air duct,
near windows, or near heating and cooling sources. Temperature
fluctuation of 1 °C or more over a 10 minutes period can affect
performance.
▲
Humidity
The relative humidity of the operating environment must be between
20 and 80%, with non-condensing and non-corrosive atmosphere. It is
recommended that your laboratory be equipped with a temperature and
humidity monitor to ensure that your laboratory is always within the
required temperature and humidity specifications.
NOTICE
Operating an iCAP RQ instrument at very low humidity
might cause the accumulation and discharge of static electricity, which
can shorten the life of electronic components. Operating the system at
high humidity might cause condensation, oxidation, and short
circuits.
▲
Altitude
The iCAP RQ instrument is designed for indoor use at an altitude of up
to 2000 m (6500 ft) above sea level. For altitudes above 2000 m, contact
Thermo Fisher Scientific.