Operating Environment
Temperature
3-2
iCAP RQ Pre-Installation Requirements Guide (P/N BRE0009927, Revision A)
Thermo Scientific
Temperature
NOTICE
After arrival, the iCAP RQ ICP-MS instrument must be stored
as packed by Thermo Fisher Scientific for at least 24 hours at an
environmental condition of 15 to 35 °C and a relative humidity of 20 to
80% before connecting it to the power supply.
▲
The mass spectrometer is designed to operate at the temperatures
specified in
NOTICE
As the laboratory temperature increases, system reliability
decreases. All electronic components generate heat while operating.
▲
Table 3-1.
Temperature requirements for iCAP RQ instruments
Specification
Value
Laboratory Temperature
15 to 35 °C
Optimum Operation Temperature
18 to 21 °C