Consumables
Gas Supply iCAP TQ/TQs ICP-MS
Thermo Scientific
iCAP Qnova Series ICP-MS Pre-Installation Requirements Guide (P/N BRE0006946, Revision A)
5-25
Oxygen
For the analysis of organic solvents, the use of oxygen may be required.
The gas requirements are detailed in
.
Pure Nitrogen or Mixtures with Helium and/or Argon
For the analysis of organic solvents, the use of nitrogen may be required.
The gas requirements are detailed in
.
Methane Mixtures with Argon
The gas requirements are detailed in
.
Table 5-24.
Additional gas requirements for oxygen
Gas Requirement
Specification
Purity
99.995% or better
Maximum Water Content
5 ppmv
Supply Rate
max. 1 L/min
Pressure Regulator Output
max. 1 bar using a two-stage
pressure regulator (200 bar to
3 bar)
NOTICE
A temperature of the spray chamber down to -10 °C will be required
for the analysis of organic solvents. This is controlled via the Peltier
cooler.
Table 5-25.
Additional gas requirements for nitrogen
Gas Requirement
Specification
Purity
99.996% or better
Maximum Water Content
5 ppmv
Supply Rate
max. 1 L/min
Pressure Regulator Output
max. 1 bar using a two-stage
pressure regulator (200 bar to
3 bar)
NOTICE
The maximum content of methane is 4.5%.
Table 5-26.
Additional gas requirements for methane
Gas Requirement
Specification
Purity
99.996% or better
Maximum Water Content
5 ppmv