Consumables
Gas Supply iCAP RQ ICP-MS
Thermo Scientific
iCAP Qnova Series ICP-MS Pre-Installation Requirements Guide (P/N BRE0006946, Revision A)
5-9
CCT Gas
The Collision Reaction Cell allows the selective removal of polyatomic
interferences.
The following services must be supplied by the customer:
•
Gas
•
Two-stage gas pressure regulator (for 1/16" capillary, 1/4" to 1/16"
adapter will be supplied with the instrument)
Helium
A CCT installation test is performed with helium gas to pressurize the
QCell™. See
for details of the helium gas requirements.
Oxygen or Mixtures with Helium (for QCell)
See
for details of the oxygen gas requirements. All mixture
ratios are permitted.
NOTICE
The CCT gas outlet should be located at a maximum distance of 2 to
3 meters to the left of the instrument. Thermo Fisher Scientific
supplies 2 x 3 meters of gas tubing with the instrument.
Table 5-5.
CCT gas requirements for helium
Gas Requirement
Specification
Purity
99.999% or better
Maximum Water Content
2 ppmv
Supply Rate
max. 10 mL/min per channel
Pressure min.
0.05 MPa
(0.5 bar)
max. 0.15 MPa (1.5 bar)
using a two-stage pressure
regulator (200 bar to 3 bar)
Table 5-6.
Gas requirements for oxygen (for QCell)
Gas Requirement
Specification
Purity
99.999% or better
Maximum Water Content
2 ppmv