iCAP Qnova Series ICP-MS Pre-Installation Requirements Guide (P/N BRE0006946, Revision B)
vii
Technical Data
iCAP RQ 2-Channel Instrument
The table summarizes the most important technical data of the
iCAP RQ 2-channel systems. See the respective chapters of the manual
for details and additional instrument properties.
Tip
If the torque plate of your instrument (see
) states
“suitable for Hydrogen use,” the following information is applicable
for your system.
Parameter
Specification
Value
Instrument Dimensions
iCAP RQ ICP-MS
Length × width × height
772 mm × 665 mm × 1102 mm
Weight
142 kg
Power Requirements
iCAP RQ ICP-MS
Nominal Voltage
200-240 V AC (50/60 Hz), single phase
Power
apparent power: 3000 VA;
effective power: 2200 W
Wire
Single-phase, 2-pole, 3-wire grounding
Fuse
15/16 A (Europe)
20 A (North America, Japan)
Protection type (acc. IEC 60529))
IP 20
Fore vacuum pump
Voltage
200-240 V (50 Hz/60 Hz)
Power
1.6 kW (50 Hz)
1.9 kW (60 Hz)
Operating Environment
Temperature
Laboratory temperature
15 to 35 °C
Optimum operation temperature
18 to 21 °C
Exhaust
Exhaust velocity (plasma)
6 to 8 m/s
Exhaust flow (plasma)
67 to 90 m³/h
Exhaust velocity (heat)
4 to 6 m/s
Exhaust flow (heat)
45 to 67m³/h
Heat generation
Instrument
350 W-520 W (50 Hz)
350 W-520 W (60 Hz)
Fore vacuum pump
900 W (50 Hz)
1100 W (60 Hz)
Humidity
20-80%, non-condensing and non-corrosive atmosphere
Cooling Water Requirements
Cooling water
Supply rate
> 5.5 L/min
Temperature
20 to 30 °C, optimum at 21 °C
Pressure
0.25 to 0.6 MPa (2.5 to 6 bar)
pH
6-8
Recommended conductivity
< 1000 μS/cm
Solid residual
< 50 μm particle size