Consumables
Gas Supply iCAP TQ/TQs ICP-MS
Thermo Scientific
iCAP Qnova Series ICP-MS Pre-Installation Requirements Guide (P/N BRE0006946, Revision A)
5-17
Gas Connections iCAP TQ/TQs ICP-MS
lists the connector types for the gases.
Labeled Components: 1=additional gas 1 inlet, 2=additional gas 2 inlet,
3=argon gas inlet, 4=H
2
inlet, 5=NH
3
inlet, 6=He inlet, 7=O
2
inlet
Figure 5-9.
Gas connection at the left of iCAP TQ/TQs ICP-MS system
Table 5-15.
Gas connections at iCAP TQ/TQs ICP-MS
Label
Connection
Tubing/ Capillary
Argon
a
a
applicable for iCAP TQ ICP-MS
push-in fitting
6 mm OD
Add1
a
push-in fitting
6 mm OD
Add2
a
push-in fitting
6 mm OD
Argon
b
b
applicable for iCAP TQs ICP-MS
bulk head union
1/4” OD
Add1
b
bulk head union
6 mm OD
Add2
b
bulk head union
6 mm OD
He
bulk head union
1/16” OD
O
2
bulk head union
1/16” OD
NH
3
bulk head union
1/16” OD
H
2
bulk head union
1/16” OD
1
2
3
7
5
6
4