Environmental requirements
Condition
Acceptable range
Installation site
Indoor use only
Electromagnetic
interference
Do not use this device in close proximity to sources of strong
electromagnetic radiation (for example, unshielded intentional
RF sources). Strong electromagnetic radiation may interfere
with the proper operation of the device.
Altitude
Between sea level and 2000 m (6500 ft.) above sea level
Operating conditions
• Humidity: 15–80% relative humidity (noncondensing)
• Temperature: 15 to 30°C (59 to 86°F)
Note:
For optimal performance, avoid rapid or extreme
fluctuations in room temperature.
Storage and transport
conditions
• Humidity: 20–80% relative humidity (noncondensing)
• Temperature: –30 to 60°C (–22 to 140°F)
Thermal output
During operation, the net thermal output, based on the
actual current draw of the instrument, is expected to be
approximately 960 W (3275 Btu/h).
Vibration
Ensure that the instrument is not adjacent to strong vibration
sources, such as a centrifuge, pump, or compressor.
Excessive vibration will affect instrument performance.
Pollution degree
The instrument has a Pollution Degree rating of II. The
instrument may only be installed in an environment that has
nonconductive pollutants such as dust particles or wood
chips. Typical environments with a Pollution Degree II rating
are laboratories and sales and commercial areas.
The noise output of the instrument is
£
55 dB(A) when running.
Other conditions
Ensure the instrument is located away from any vents
that could expel particulate material onto the instrument
components.
Avoid placing the instrument and computer adjacent to
heaters, cooling ducts, or in direct sunlight.
Instrument specifications
Environmental requirements
iBright
™
CL750 Imaging System User Guide
45