Run instrument diagnostics
Instrument functions can be checked by a user to confirm proper hardware operating
conditions.
1.
Touch .
2.
Touch
Service Tools
.
3.
Touch
Instrument diagnostic
.
4.
Touch
Start test
.
The instrument will check the instrument hardware for correct operation and
settings. The test will take ~10 minutes to complete. The results of the test will
display once complete. Detailed results can be exported for further review.
Note:
Ensure the drawer is not blocked. The drawer will open as part of the
testing. (Not applicable to iBright
™
CL750 model.)
Clean the instrument
Repeated instrument use can cause spots and smudges on the sample stage, which
can then decrease image quality. Clean the sample stage as needed.
Materials required
• Safety glasses
• Powder-free gloves
• Tissue, lint-free
• Deionized water
• Ethanol, 70% solution
Note:
Avoid the use of detergents.
1.
Open the drawer to expose the sample stage.
2.
Lightly spray the glass surface with deionized water or a 70% ethanol solution.
3.
Wipe the surface with a lint-free tissue until sufficiently clean.
4.
Close the drawer and operate the instrument as normal.
Maintain the instrument
Run instrument diagnostics
30
iBright
™
CL750 Imaging System User Guide