Table 7 Other options
(continued)
Suboption
Detail
Smart exposure
Predicts optimal exposure for minimizing pixel saturation
and maximizing dynamic range for a specific sample.
Renders a preview of how the image will appear at the
recommended exposure time.
Band excision
(Nucleic
Acid Gels mode)
Provides toggle button to turn on/off the LED light to allow
for band excision on nucleic acid gels.
Membrane overlay
(Chemi Blots mode)
Allows visible membrane overlay to see any visible
prestained markers that aid in identifying molecular
weights for unknown samples.
2UP view
(Universal Mode)
After a
Smart Exposure
or
Capture
, images are
displayed by default in a
2UP view
.
• Top image displays individual channels in grayscale.
Switch between individual channels by touching the
channel you want to display and edit. After a dye is
selected, that channel image is displayed in the top
view and becomes editable so the user can increase
or decrease the exposure time dial and see in real
time the effect on the image preview.
• Bottom image displays the color composite with
channels displayed as different false colors overlaid
on a black background. Individual channels can
be toggled on and off in the composite image by
touching the eyeball at the left of the appropriate
channel name.
• For the
Visible
channel color composite, the
signal is false
‑
colored overlaid on a white
background. Example: This channel is used to display
Coomassie
‑
stained protein gels.
1UP view
(Universal Blot
mode)
After a
Smart Exposure
or
Capture
, images can also be
displayed in a
1UP view
.
Image displays the grayscale and false-color composite
images by toggling
Gray
or
Color
. User can switch
between individual channels by touching the dye name
you want to display. Individual channels can be toggled on
and off in the composite image by touching
on the left
of the appropriate dye name.
Note:
Functionality for toggling between individual
channels in grayscale view or toggling on and off individual
channels in the false-color composite view is identical to
the
2UP view
.
Auto-enhance
Removes the channel level background to enhance the
displayed image.
Operate the iBright
™
imager
Options available in modes
iBright
™
CL750 Imaging System User Guide
19