Instrument specifications
Instrument dimensions and specifications
Specification
iBright
™
Imager
Dimensions (L × W × H)
68.0 × 38.2 × 59.8 cm
Weight
52 kg (115 lbs)
Instrument clearance (Back)
15.25 cm (6 in.)
Electrical requirements
WARNING!
For safety, the power outlet used for powering the instrument must
be accessible at all times. In case of emergency, you must be able to immediately
disconnect the main power supply to the instrument. Allow adequate space between
the wall and the equipment so the power cord can be disconnected in case of
emergency.
• Electric receptacle with grounding capability
• Maximum power dissipation: ~960 W (not including computer and monitor)
• Mains AC line voltage tolerances must be up to ±10 percent of nominal voltage
Device
Rated voltage
Circuit
required
Rated
frequency
Rated
power
iBright
™
Imager
100–240 ±10%
VAC
[1]
10 A
50/60 Hz
410 W
[1]
If the supplied power fluctuates beyond the rated voltage, a power line regulator may be required. High or
low voltages can adversely affect the electronic components of the instrument.
44
iBright
™
CL750 Imaging System User Guide