Thermo Fisher Scientific CryoExtra CE8100 Series
Appendix | 73
LOTBS?
Temp B Low Alarm Status Query
Input: LOTBS?
Returned:
0 or 1
Comments:
Used to query the Temp B Low Alarm status
where 0 = off and 1 = on.
LOTBM?
Temp B Low Alarm Mask Query
Input: LOTBM?
Returned:
0 or 1
Comments:
Used to query the Temp B Low Alarm Mask
status where 0 = disabled and 1 = enabled. Default setting is
1 = enabled.
TEMPA?
Temp A Current Data Query
Input: TEMPA?
Returned:
±XXX.X
Comments:
Used to query the current Temp A sensor
reading.
TEMPB?
Temp B Current Data Query
Input: TEMPB?
Returned:
±XXX.X
Comments:
Used to query the current Temp B sensor
reading.
TUNI
Set Temp Units
Input: TUNI C, TUNI F or TUNI K
Returned:
No response
Comments:
Used to set the Temp units where C = Celsius,
F = Fahrenheit and K = Kelvin. Default setting is degrees
Celsius.
TUNI?
Temp Units Query
Input: TUNI?
Returned:
C, F or K
Comments:
Used to query the Temp units where
C = Celsius, F = Fahrenheit and K = Kelvin. Default setting is
degrees Celsius.
Level Commands
BPFIL?
Bypass Status Query
Input: BPFIL?
Returned: 0 or 1
Comments:
Used to query the current hot gas bypass status
where 0 = not bypassing and 1 = bypassing.
FILAS?
Fill Time Alarm Status Query
Input: FILAS?
Returned:
0 or 1
Comments:
Used to query the Fill Time Alarm Status where
0 = off and 1 = on.
FILL
Set Fill Status
Input: FILL 0 or FILL 1
Returned:
No response
Comments:
Used to set the current Fill Status
where 0 = Stop Fill and 1 = Start Fill.
FILL?
Fill Status Query
Input: FILL?
Returned:
0 or 1
Comments:
Used to query the current Fill Status
where 0 = not filling and 1 = filling.
FILLM?
Auto Fill Status Query
Input: FILLM?
Returned:
0 or 1
Comments:
Used to query the Auto Fill Status
where 0 = disabled and 1 = enabled.
FILT
Set Max Fill Time
Input: FILT XXX
Returned:
No response
Comments:
Used to set the maximum fill time in minutes
from 30 to 240 minutes.
FILT?
Max Fill Time Query
Input: FILT?
Returned:
XXX
Comments:
Used to query the maximum fill time. Value
returned in minutes.
FILTIM?
Fill Duration Query
Input: FILTIM?
Returned:
XXX
Comments:
Used to query the duration of the current fill.
Value returned in seconds.
HFIL
Set High Level Setpoint
Input: HFIL XXX.X
Returned:
No response
Comments:
Used to set the High Level Setpoint. Input value
uses the current level units, either inches or millimeters. There
must be at least a 0.5 inch interval between each level setting
and alarm.