72 | Appendix
Thermo Fisher Scientific CryoExtra CE8100 Series
HITA?
Temp A High Alarm Query
Input: HITA?
Returned:
±XXX.X
Comments:
Used to query the current Temp A High Alarm
value.
HITAS?
Temp A High Alarm Status Query
Input: HITAS?
Returned:
0 or 1
Comments:
Used to query the Temp A High Alarm status
where 0 = Off and 1 = On.
HITAM
Set Temp A High Alarm Mask
Input: HITAM 0 or HITAM 1
Returned:
No response
Comments:
Used to enable or disable the Temp A High
Alarm where 0 = disabled and 1 = enabled. Default setting is
1 = enabled.
HITAM?
Temp A High Alarm Mask Query
Input: HITAM?
Returned:
0 or 1
Comments:
Used to query the Temp A High Alarm Mask
status where 0 = disabled and 1 = enabled. Default setting is
1 = enabled.
HITB
Set Temp B High Alarm
Input: HITB ±XXX.X
Returned:
No response
Comments:
Used to set the Temp B High Alarm value.
HITB?
Temp B High Alarm Query
Input: HITB?
Returned:
±XXX.X
Comments:
Used to query the current Temp B High Alarm
value.
HITBS?
Temp B High Alarm Status Query
Input: HITBS?
Returned:
0 or 1
Comments:
Used to query the Temp B High Alarm status
where 0 = Off and 1 = On.
HITBM
Set Temp B High Alarm Mask
Input: HITBM 0 or HITBM 1
Returned:
No response
Comments:
Used to enable or disable the Temp B High
Alarm where 0 = disabled and 1 = enabled. Default setting is
1 = enabled.
HITBM?
Temp B High Alarm Mask Query
Input: HITBM?
Returned:
0 or 1
Comments:
Used to query the Temp B High Alarm Mask
status where 0 = disabled and 1 = enabled. Default setting is
1 = enabled.
LOTA
Set Temp A Low Alarm
Input: LOTA ±XXX.X
Returned:
No response
Comments:
Used to set the Temp A Low Alarm value.
LOTA?
Temp A Low Alarm Query
Input: LOTA?
Returned:
±XXX.X
Comments:
Used to query the Temp A Low Alarm value.
LOTAS?
Temp A Low Alarm Status Query
Input: LOTAS?
Returned:
0 or 1
Comments:
Used to query the Temp A Low Alarm status
where 0 = off and 1 = on.
LOTAM
Set Temp A Low Alarm Mask
Input: LOTAM 0 or LOTAM 1
Returned:
No response
Comments:
Used to enable or disable the Temp A Low
Alarm where 0 = disabled and 1 = enabled. Default setting is
1 = enabled.
LOTAM?
Temp A Low Alarm Mask Query
Input: LOTAM?
Returned:
0 or 1
Comments:
Used to query the Temp A Low Alarm Mask
status where 0 = disabled and 1 = enabled. Default setting is
1 = enabled.
LOTB
Set Temp B Low Alarm
Inp
ut: LOTB ±XXX.X
Returned:
No response
Comments:
Used to set the Temp B Low Alarm value.
LOTB?
Temp B Low Alarm Query
Input: LOTB?
Returned:
±XXX.X
Comments:
Used to query the Temp B Low Alarm value.