Thermo Fisher Scientific CryoExtra CE8100 Series
Operation | 35
Note:
It is recommended to download complete data from
the Controller prior to performing a restore to defaults.
Restoring to defaults will erase previous stored data and it
cannot be retrieved once the restore function is performed
.
Note:
Security Level 3 or higher is required to reset to factory
default settings.
1.
Press “SETUP”
Controller will prompt for a
password. A flashing cursor on the
entry screen will make it clear
which digit is being changed. Use
the “
” keys to scroll to the appropriate number. Press
“ENTER” to advance the cursor to the next position.
2.
Press “ENTER”
The display will read “Temperature
Menus”.
3.
Press “SETUP”
Until the display reads “Advanced
Settings”.
4.
Press “ENTER”
The display will read “Advanced Fill
Menus”.
5.
Press “SETUP”
Until the display reads “Restore All
Defaults”.
6.
Press “ENTER”
The display will read “Confirm
Restore”. Use the “
” keys to
change to “YES” and press
“ENTER” to save.
Controller Default Settings
Restart Controller
The Controller can be restarted in the Advanced Settings
Menu. Restarting the Controller and cycling the power are
both safe ways to reboot the Controller.
Note:
Security Level 3 or higher is required to restart the
Controller from the Advanced Settings Menu.
1.
Press “SETUP”
Controller will prompt for a
password. A flashing cursor on the
entry screen will make it clear
which digit is being changed.
Higher User Level
Required use
to
ENTER Password 0000
Press ENTER for
Temperature Menus or
press SETUP for next
menu
Press ENTER for
Advanced Settings or
press SETUP for next
menu
Press ENTER for
Advanced Fill Menus
or press SETUP for next
menu
Press ENTER to Restore
All Defaults or press
SETUP for next menu
Confirm Restore NO
Use
to
adjust
Press ENTER to save
Parameter
Default Setting
High Temp Alarm
-110.0°C / -166.0°F / 163.2 K
Low Temp Alarm
-200.0°C / -327.6°F / 73.4 K
High Level Alarm
8.0 in / 205 mm
High Level Setpoint
7.0 in / 180 mm
Low Level Setpoint
5.0 in / 125 mm
Low Level Alarm
4.0 in / 100 mm
Defined 100%
8.0 in / 205 mm
Defined 0%
4.0 in / 100 mm
Level Offset
+1.3 in / +35 mm
Auto Fill
Enabled
Maximum Fill Time
60 minutes
Temperature Display
Units
°C
Liquid Level Display
Units
Inches
Hot Gas Bypass MENU
DISABLED
Hot Gas Bypass Temp
Setpoint
-70°C
Hot Gas Bypass Alarm
Time Delay
5 minutes
COM 1 Type
ASCII
COM 2 Type
Disabled
Event Log Interval
240 minutes
Global Password
3 4 5 6
Higher User Level
Required use
to
ENTER Password 0000