Service Manual BB 150
Index
Name
Date
Document No.
Page
File
50078923_04_CN
50 079 823
17/ 17
21 Unit configuration
FL position
FL value range
Description
1
0 = BB 150 / RB 150/ BB 15
1 = BB 24 / RB 240
Configure unit size.
2
0 = VA
1 = CU
Configure interior fittings.
3
0 = 230 Volt
1 = 120 Volt
Configure mains voltage.
4
0 = Not installed
1 = Installed
Configure gas diaphragm.
5
0 = No O
2
Configure
O
2
.
6
0 = Not installed
Configure flask rotator.
7
0 = Standard TCD
Configure installed measuring cell.
8
0 = Not installed
Configure water level monitor.
9
1 = Not installed
Configure low humidity option.
10
0 = Not installed
Gas guard CO2.
11
0 = Not installed
Gas guard O2/N2.
12
1 = not active
Contra Con Routine
13
0
- without function -
14
0
- without function -
15
0
- without function -
16
0
- without function -
17
0
- without function -
18
0
- without function -
19
0
- without function -
20
0 / 1
Start write process.
21
0 / 1
Security bit for position 1-10
For initial configuration only!
Using this function will overwrite all unit adjustments on the main board!
When FL 601 is called up, the system jumps to FL 21.