11/ 11
Table 3/1: Function levels (FL)
CAUTION!
Function levels are for factory settings only, or are automatically determined.
The settings may only be modified after consultation with the factory.
Levels
1
Adjustments for regulating range 1 (incubation mode, 0 - 55 °C
)
Sublevel
[CO
2
display]
Sublevel value range
[temperature display]
Description
1
0 = (+) / -0 = (-)
Temperature offset value sign
2
0.0 ..... 99.9
Offset value temperature, in °C
3
0. ..... 1
Temperature amplification, pre-decimal
places
4
000 ..... 999
Temperature amplification, decimal places
5
0. ..... 1
cal factor, pre-decimal places
6
800.....999
/
000.....200 cal
factor, decimal places
7
0 = (+) / -0 = (-)
auto-start offset value sign (NV RAM on
the sensor board)
8
0.0 ..... 99.9
auto-start offset, in % CO
2
(measuring
cell, NV RAM)
9
0. ..... 1
CO
2
amplification, pre-decimal places
(measuring cell, NV RAM)
10
000 ..... 999
CO
2
amplification, decimal places
(measuring cell, NV RAM)
11
0 = (+) / -0 = (-)
cal value sign
12
0.0 ..... 99.9
cal value, in % CO
2
The items identified by "-cal-" are automatically modified by the adjustment routine
initiated with the