Name
Date
Designation
Prepared
JF 05.07.13
Service Manual BB 150
Index
Name
Date
Document No.
Page
File
50078923_04_CN
50 079 823
12/ 12
2
Not reserved
FL position
FL value range
Description
*
3
Outputs, manual control (main PCB)
FL position
FL value range
Description
1
0 / 1
Heater actuator, unit body: X5...X8
2
0 / 1
Heater actuator, unit door: X3
3
0 / 1
Heater actuator, base: X10
4
0 / 1
Actuator, CO
2
valve: JP5
5
0 / 1
Actuator, common alarm: X11
6
0 / 1
Actuator, horn
7
0 / 1
Actuator, fan
8
0 / 1
Actuators remain active as long as the
cal
key is pressed.
9
0 / 1
O
2
valve switch
The output actuators can be manually activated on this function level. The switch state
(1) can be reset by:
- Manually returning it to 0, or;
- Performing a mains reset.