Service Manual BB 150
Index
Name
Date
Document No.
Page
File
50078923_04_CN
50 079 823
28/ 28
Decontamination status:
You can also output the precise phase of the decontamination routine that the unit is
currently going through. The phase is shown on the CO
2
display.
Simultaneously press and hold down the –90°C- and -i- keys for approx. 5 seconds:
[ S - - ]
Warm-up phase
[ - S - ]
Decontamination phase
[ - H - ]
Condensation phase
[ - - S ]
Cool-down phase
[ - - H ]
Reheat phase
- Circuit diagram:
See circuit diagrams in Chapter 3.
Technical specifications - same as for the incubation mode:
Nominal voltage: 230 VAC
Power consumption:
0.63
kW