Name
Date
Designation
Prepared
JF 05.07.13
Service Manual BB 150
Index
Name
Date
Document No.
Page
File
50078923_04_CN
50 079 823
14/ 14
8
Door heating factor, on/off
FL position
FL value range
Description
1
PH0
PH1
Door heating factor OFF
Door heating factor ON
2
0.0 .... 3.0
Door heating factor
Factory setting: Door factor ON = PH1. May not be changed!
When replacing the main board, the door heating factor for the model in
question must be checked in function level 8, position 2. Readjust if
necessary:
Units with stainless steel interior fittings: 1.4
Units with copper interior fittings:
2.2
9
Display brightness
FL position
FL value range
Description
1
1 .... 15
Display brightness
Factory setting: Brightness stage 8
10 Not reserved
FL position
FL value range
Description