CO
2
Incubator BB 150 Operating Instructions
27
5.
Start-up
5.5
Power supply connection
WARNING – Electric shock!
Contact with current-carrying components may
cause a lethal electric shock.
Before connecting the device to the power
supply, check plug and connection line for
damage.
Do not use damaged components when con-
necting the device to the power supply!
The device must be connected only to a correctly installed and
grounded power supply source:
• Fusing
T 16 A
• Circuit breaker
G 16
Connection to the power supply system:
1. Before connecting the device to the power supply, check
to see if the voltage of the power supply corresponds with
the specifications on the nameplate at the front of the de-
vice. If the ratings given for voltage (V) and current (A) are
not correct, the device must not be connected to the power
supply.
2.
Fig. 15:
Connect the connector for non-heating appliances
[2] to the socket [1] at the control box of the device.
3. Connect the grounding plug [3] of the power supply cable
to a correctly grounded and fused socket.
4. Make sure the power supply cable is not subjected to tensile
or compressive force.
Fig. 15
Power supply connection