Chapter |
Operation
Confidential, limited rights
PN AA83865 | Revision A | 25-Oct-2019
Instruction Notice, ARL EQUINOX 5000-5500
Page
38
7
Operation
Before powering the instrument ON, double-check electrical, gas and water distributions. Also
check if the cooling fluid tank is full and all protection panels and covers are mounted.
7.1 Start Up
1.
Turn the main breaker located on the rear panel to the "I-ON" position.
2.
If the instrument is equipped with circuit breakers, make sure they are in ON position as
well.
3.
Check that the emergency stop button is released.
4.
When the white button marked "Power ON" located on the front panel is illuminated, push
on it.
5.
After powering the instrument ON, the red light of the "Power OFF" button turns ON.
6.
If the instrument is connected to an external cooling circuit, make sure the source of the
cooling water is opened, with appropriate characteristics.
7.
After several tens of seconds the system will be ready for operation.
Note
The ARL EQUINOX 5000/5500 X-Ray diffractometer driving operations are made by
Symphonix software. Please, read the Symphonix user guide for further inform ation
on how to use the instrument. The Symphonix user's guide is available in PDF file
format in the folder c:/inel/Documentation/SymphoniX user guide.pdf