Chapter |
Connection Specifications
Confidential, limited rights
PN AA83865 | Revision A | 25-Oct-2019
Instruction Notice, ARL EQUINOX 5000-5500
Page
28
5
Connection Specifications
5.1 Electrical Connection
ARL EQUINOX 5000/5500 X-Ray diffractometers have an Ingress Protection rating of IP20.
This means that they are protected against the ingress of solid particles with a diameter over
12 mm (e.g.: a finger). ARL EQUINOX 5000/5500 X-Ray diffractometers do not have any
protection against the ingress of liquids.
All connections to the instrument have to comply with class SELV (Safety Extra Low Voltage)
according to norm EN 61010-1:2010.
Equipment
Voltage [V]
Frequency [Hz]
Phase
Power [W]
Max consumption
[A]
ARL EQUINOX
230
10%
50/60
1
-
32
Water chiller
110/230
50/60
1
-
15
Computer
100/240
50/60
1
250
-
Screen
100/240
50/60
1
-
1
The ARL EQUINOX 5000/5500 complies with overvoltage category II (OVCII).
The ARL EQUINOX 5000/5500 complies with insulation class 1.
WARNING!
GROUND CONNECTION: To guarantee a proper grounding of the instrument, a
ground wire of 6mm² cross-section has to be connected to the ground terminal