Thermo Scientific AquaSensors
™
DataStick
™
Ethernet Communications User Guide
72
Bank 1000 contains registers that are used for informational purposes only.
Table 5: Register bank 1000; status registers.
Register
Number
Description
Data
Type
Access
DataStick
Command
Comment Example
Value
Variable
Class
41,001
N10:01
Communications
Status
integer
R N/A
Status
41,002
N10:02
Calibration
Status
integer
R CALSTATUS
Chars:
2;
Offset: 6
1 Status
41,003
N10:03
Sensor Memory
Status
integer
R GSTATUS
Chars:
1;
Offset: 0
1 Status
41,004
N10:04
Configuration
Memory Status
integer
R GSTATUS
Chars:
1;
Offset: 2
1 Status
41,005
N10:05
Calibration
Memory Status
integer
R GSTATUS
Chars:
1;
Offset: 4
1 Status
41,006
N10:06
Run Status
integer
R
GSTATUS
Chars: 1;
Offset: 6
1 Status
All of the registers in Bank 2000 contain floating-point values. Use these registers if the
other devices on the network support floating-point values. If the other devices on the
network do not support floating-point values, corresponding integer values can be found
in Bank 3000.
If the product is not attached to a DataStick body, the Sensor Value (float) and the
Temperature Value (float) shown in Table 6 will read “Not a Number” (NaN).
Table 6: Register bank 2000; floating-point registers.
Register
Number
Description
Data
Type
Access
DataStick
Command
Comment
Example Value
Variable
Class
42,001
N20:01
Sensor Value
float
R
GSNSR
7.04
Process
Value
42,002
N20:02
"
42,003
N20:03
Temperature
Value
float R GTEMP
25.1
Process
Value
42,004
N20:04
"
42,005
N20:05
Calibrate
Sensor 1-
Point Sample
Value
float R/W CALS1PS
7.02
Calibration
42,006
N20:06
"
42,007
N20:07
Calibrate
Sensor 2-
Point Sample
Value
float R/W CALS2PS
10.04
Calibration
42,008
N20:08
"
42,009
N20:09
Calibrate
Temperature
1-Point
float R/W CALST1PS
25.3
Calibration