Thermo Scientific AquaSensors
™
DataStick
™
Ethernet Communications User Guide
81
Register
Number
Description
Data
Type
Access
DataStick
Command
Comment
Example
Value
Variable
Class
45,022
N50:22
Sensor
Type
integer R GSTYPE Chars:
2
1 Information
45,023
N50:23
Calibration
Type
integer R CALSTATUS
Chars:
2;
Offset: 3
4 Information
45,024
N50:24
Comms
Adapter
Code
Version
integer
R N/A V3.03
0x0303
Information
45,025
N50:25
45,026
N50:26
Location01 ASCII[2] R/W
N/A
Also...
"TA"
Information
45,027
N50:27
Location02 ASCII[2] R/W
N/A
readable… "NK"
Information
45,028
N50:28
Location03 ASCII[2] R/W
N/A
by…
"#3"
Information
45,029
N50:29
Location04 ASCII[2] R/W
N/A
Discovery…
"BL"
Information
45,030
N50:30
Location05 ASCII[2] R/W
N/A
Tool.
"DG"
Information
45,031
N50:31
Location06 ASCII[2] R/W
N/A
“
"#8"
Information
45,032
N50:32
Location07
ASCII[2]
R/W N/A
“
Information
45,033
N50:33
Location08
ASCII[2]
R/W N/A
“
Information
45,034
N50:34
Data Valid
integer
R
N/A
Information
45,035
N50:35
Pending
Writes
integer R
N/A
Information
45,036
N50:36
Read Error
Count
integer
R/W
N/A
Resets to 0
on write
Information
45,037
N50:37
Write Error
Count
integer
R/W
N/A
Resets to 0
on write
Information
45,038
N50:38
Timeout
Count
integer
R/W
N/A
Resets to 0
on write
Information
When the Ethernet option is present in the AV38 Local Display, the register in Table 10
can be used to change the address of the DataStick with which the AV38 is
communicating. In this way, a device on the Ethernet network can access a DataStick
that’s connected to the AV38.
Table 10: Register bank 6000; AV38 configuration register.
Register
Number
Description
Data
Type
Access
DataStick
Command
Comment
Example
Value
Variable
Class
46,001
N60:01
DataStick
Slave
Address
integer R/W GDSA/
SDSA
123
AV38
Configuration