Thermo Scientific AquaSensors
™
DataStick
™
Ethernet Communications User Guide
39
To perform a 1-point sample calibration on the Sensor or the Temperature measurement,
enter the desired value in the Sensor or Temperature field and click the corresponding Set
button. The calibration procedure can be monitored by returning to the Home Page and
observing the Sensor or Temperature Value, and the Last Cal Type and Calibration Status
fields. Calibration Status will indicate Cal OK if the procedure completed successfully.
Clicking the Cancel button instead of the Set button will return the 1-Point Sample
Calibration value to its previous value.
5.3 Configuration Page
The configuration edit page can be accessed by clicking on any of the Measurement Units
or Configuration Values on the Home Page as shown in Figure 47.
Figure 47: Accessing the Configuration Edit Page.
Click on any of the
Measurement Units
or Configuration
Values to access the
Configuration Edit
Page