Thermo Scientific AquaSensors
™
AquaChlor
™
Free Chlorine Monitor User Guide
24
≡
Main Menu
►
Reset AV38
▲
▼
►
◄
ESC
ENTER
MENU
To return to the measure screen press ENTER when in the EXIT screen. Pressing ESC from any screen
will take the user up one level. In this example ESC will also take the user to the measure screen. Since
this screen is at the bottom of the main menu list, the down arrow will not work here. The up arrow can be
used to go back through the list.
≡
Main Menu
◄
Exit
MENU
ESC
ENTER
◄
►
▼
▲
5.3.
1-Point (Span) Calibration
To calibrate the span for the Free Chlorine System:
1.
Draw a sample with a syringe from the process water in the sample chamber and check the
concentration with a lab meter. For the most accurate calibration the reading should be above 1 ppm.
2. Perform a 1-point sample calibration using the reading from the lab meter.
The complete calibrate menu for free chlorine is shown.
≡
Calibrate
►
1-Point Sample
►
Temperature
◄
Exit
▲
►
▼
◄
MENU
ESC
ENTER
From the measure screen, press MENU.