Thermo Scientific AquaSensors
™
AquaChlor
™
Free Chlorine Monitor User Guide
27
5.4.
Temperature Calibration
Press the ENTER key from the TEMPERATURE calibration menu to initiate the calibration process.
When the temperature calibration procedure is started, the analog output is placed into hold mode. The
user is prompted to prepare the sensor. The sensor value is dynamically updated during this step.
When the sensor is ready, the ENTER key is pressed and the user is presented with a calibration value for
editing. Please note that it is best to calibrate temperature when the sample chamber is full of flowing
process water.
≡
Calibrate
►
Temperature
MENU
ESC
ENTER
◄
►
▼
▲
Temperature?
( 25.3) °C
MENU
ESC
ENTER
◄
►
▼
▲
25.3 °C
Sensor ready?
MENU
ESC
ENTER
◄
►
▼
▲